Portable Leeb Hardness Tester ML108

Brief Introductions:

Modern Plastic cover,offor dable, variant of the ML108 with additional  features.Easy operationg due to intuitive menu system, Fast, support optional built-out mini-printer.
 

Instrument Features:

             Large memory for 500 groups  tests
       Auto-recognition of connected Impact Device
Auto-recognition of Impact Device test direction
 Statistical data and upper-lower limit setting
RS-232 interface (to hyperterminal or dataview software)
Time and date setting; auto-clock
Back-light LCD
Direct display of hardness scales Rockwell HRB, HRC, Vickers
HV, Brinell HB, Shore HS, Leebs HL
 Conversion to tensile strength σb (U.T.S.)
For all metallic materials
Impact device provides testing at any angle, even upside down
Integral thermal printer
High accuracy ±6HLD and conforms to ASTM A 956
 Six impact devices are available for special applications

 

Technical Specifications

 

Specifications
ML108
Order#
852-108
Hardness parameter
HL,HRC ,HRB,HRA,HV,HB,HS
Tensile strength U.T.S. range(steel only)
ób from 374 to 2652
Accuracy
Within±6HLD
Functions
Auto-recognition of connected probe and probe test direction, auto-conversion to other hardness scales
Memory
500 test maximum
Data-output
RS-232 to hyper terminal (MS) and data view TH-160
Statistics
Average value, min-max ,upper-lower limits
Impact device
D (standard)
Optional impact devices
DC/D+15/DL/G/C/E
Max. hardness value
900HV-1000HV
Work piece radius
(convex/concave)
Rmin=50mm (with support ring Rmin=10mm)
Min. Work piece weight
2~5kg on stable support
0.05~2kg with compact coupling
Except C and G impact device
Work piece min. thickness coupled
5mm,expect with impact device G:10mm,C:1mm
Work piece min. case hardened depth
0.08mm ,expect C impact device 0.2mm and G impact device 1.2mm
Power
AA3V
Operating temperature
0 to 40 ℃
Overall dimensions
131mmx75mmx31mm
weight
200g(including impact device and printer)

 

Accuracy and repeatability of displayed value: 

No.
Type of impact device
Oreder#
Hardness value of Leeb standard hardness block
Erro of displyed value
Reapeatability
1
D
852-418
760±30HLD
530±40HLD
±6 HLD
±10 HLD
6HLD
10HLD
2
DC
852-428
760±30HLD
530±40HLD
±6 HLDC
±10 HLDC
6HLD
10HLD
3
DL
852-438
878±30HLD
736±40HLD
±12 HLDL
12HLDL
4
D+15
852-458
766±30HLD
544±40HLD
±12 HLD+15
12HLD+15
5
G
852-478
590±40HLD
500±40HLD
±12 HLG
12HLG
6
E
852-468
725±30HLD
508±40HLD
±12 HLE
12HLE
7
C
852-448
822±30HLD
590±40HLD
±12 HLC
12HLC

 

Standard Deliveries:
Main unit with impact device type D
Test block with HLD-value
Cleaning brush
RS-232 Data cable
SINOWON certificate
Operation Manual
Carrying case
 
Optional Accessories:
Optional Mini-printer
Optional Support Ring